The camera system consists of a camera host, a light source, a pneumoperitoneum and various connecting wires. Today, we have sorted out six common faults and treatment methods of cold light source for your reference.
P1 Time-consuming Meter Flash
Causes of failure:
Xenon Lamp Over 500 Operating Hours
Fault handling:
Replacement of Xenon Lamp and Replacement of Time-consuming Meter
P2 device is not running, LED light on power switch is not on
Causes of failure:
1. Power switch not turned on
2. Wire damage of power supply
3. No Voltage
Fault handling:
1. Turn on the power switch
2. Replacement of power fuse
3. Check the power supply voltage
Closing of P3 devices during operation
Causes of failure:
Overheated
Fault handling:
Eliminate the factors that hinder heat dissipation (covering, device, blocked fan, etc.)
P4 can't light xenon lamp and can't hear trigger pulse sound
Causes of failure:
Ballast damage
Fault handling:
Contact Maintenance Department
P5 can't light xenon lamp and can't hear trigger pulse sound
Causes of failure:
1. Instrument overheating (overheating circuit to start)
2. Xenon lamp damage
Fault handling:
1. Let the equipment cool down
2. Replacement of xenon lamps
P6 Brightness Insufficient
Causes of failure:
1. Incorrect installation of xenon lamp
2. Xenon lamp aging
3. Cable damage
Fault handling:
1. Check the correct installation
2. Replacement of xenon lamps
3. Use undamaged optical cables